Chemiluminescence Immunoassay Analyzer
Chemiluminescence Immunoassay Analyzer uses a modular design with a single module capable
of a throughput of up to 320 T/H. The modular design allows a combination of 1 to 4 modules
offering a maximum throughput of 1280T/H with each module capable of independent testing
without any interference between modules.
Avidin-Acridinium Ester Chemiluminescence Detection Platform
·Internationally recognized mainstream technology.
·High sensitivity, low reactant background with accurate test results.
·Higher stability for reagents with storage at 2-8 °C
Efficient Testing
·Sample Unit
·Basket loading type can easily handle large capacity loader
·620 sample positions, including 20 STAT positions
·Supporting random placement bi-directionally on sample rack for better convenience
Incubation Disk
·180 incubation positions, maintaining a constant temperature at 37°±0.3°C to maximally be close
to human body temperature.
·Time for the first result within 15 mins.
Track Unit
3 independent tracks allows the insusceptibility transportation of samples
Reagent Disk
·36 reagent positions, refrigerated at 2-8°C
·Continuous loading
·Non-contact mixing method
“Cuvette System:
·Unlimited loading of cuvettes with super large quantity of 1500 pcs loading at one time.
·Non-stop disorderly loading of reaction cuvettes.
·Non-stop loading and auto-diluting for washing buffer
Technical Parameters:
| Specification | Value |
| Technology principle | Magnetic chemiluminescence |
| Detection Platform | Avidin-Acridinium Ester Chemiluminescence |
| Throughput | 180 T/H |
| Reagent position | 25 positions |
| Reagent storage temperature | 2~8C |
| Sample position | 90 positions |
| Barcode reader | Internal |
| Sampling technology | Liquid level detection |
| Interface | Network interface |
| Connect LIS/HIS system | Able to connect LIS/HIS |
| Weight | About 200kg |
| Power consumption | About 1000VA |
| Dimension | 1230mm×757mm×554mm(L×W×H) |
| Power supply | 220±22VAC/110±11VAC |