Description
- Holographic concave flat field grating with rear spectrophotometry and cluster optical path to achieve micro volume detection of reaction solution.
- Polished probe with functions of clot detection, collision protection and automatic resetting.
- Microprocessor-controlled, fully automated biochemistry analyzer
- User-programmable with built-in barcode reader
- Sample types: Serum, Plasma, Urine Body Fluids, etc.
- Automatic re-run with onboard sample dilution
- Automatic sample quality analysis for Lipemia, Hemolysis, and Icterus
- Throughput: Up to 400 tests
- >80 tests on panel, including ISE
- Minimum >=80 sample/QC positions with continuous sample loading
- Spectral range: 320–690 nm
- Diffraction grating optics
- Sample volume: 2–35 µL
- Onboard reagent refrigeration: 2–8°C, compatible with reagent kits
- Facility to display actual reagent temperature
- Assay temperature control at 37°C
- Self-calibration against known standards
- Automatic calibration curves and result calculation
- Storage of QC results
- Independent STAT and reflex testing capability
- Touch-screen display monitor
- Data entry via keyboard, barcode reader, and LIS
- Automatic flagging of results outside user-defined limits
- Data storage capacity: 100,000 test results
- CD/USB storage system
- RS-232 interface for online computer connectivity
- Bubble detection and liquid level sensing system
- Operates on 220V, 50 Hz
- Reliable refrigerated system for reagents and sample positions.
- The on-board hemolysis function, realize automatically whole blood test for HbA1c
- Auto-dilution and Auto-rerun function
- Bi-directional LIS/HIS system to realize remote operation, and minimal maintenance.
EXCELLENT PERFORMANCE
- Automatically cuvette checkup ensure the cuvette clean
- Collision Protection
SAVE TIME AND COST
- Optical diameter 5mm
- Minimum reaction volume 120pL
HIGH RELIABILITY
- Large Absorbance Linear Range: Up to 3.6Abs
- High traceability with original reagents, calibrators and Controls
THE ON-BOARD HEMOLYSIS FUNCTION
- NGSP certification of whole blood latex-agglutination method on analyzer
- Whole blood testing function for HbA1c to avoid artificial error
- Not necessary for centrifugation
- Automatically hemolysis function, easy operation and standardization for HbA1c test
ACCURATE RESULT
- 340-800nm 12 wavelengths
- Digital liquid level detection high sensitivity avoids bubbles’ interference
- Reagent inventory management, automatically calculate remaining volume and remaining number
CHARACTERISTICS
| Category | Specification |
| Operation | automatic |
| Configuration | floor-standing |
| Technology | with ISE |
| Sample type | blood, whole blood |
| Options | with STAT port |
| Throughput | 400 p/h, 800 p/h |
| Sample volume | Max.: 0.35 ml (0.011835 US fl oz)
Min.: 0.0015 ml (0.000051 US fl oz) |
| Number of sample positions | 115 unit |
| Reagent volume | Max.: 0.35 ml (0.01 US fl oz)
Min.: 0.02 ml (0 US fl oz) |
| Number of reagent positions | 112 unit |
| Weight | 300 kg (661.4 lb) |
| Length | 1,060 mm (41.7 in) |
| Width | 790 mm (31.1 in) |
| Height | 1,150 mm (45.3 in) |
Accessories
- Built-in thermal printer or external laser printer
- Imported online sine-wave UPS with minimum 30-minute battery backup
- Compatible RO water purification system for analyzer operation

Reviews