MBI-B32
Intensive Care Baby Incubator
Features:
- LCD display.
- Double control modes of bed temperature and skin temperature.
- Double-layer thermostatic cover, automatic air curtain device.
- Humidity control function.
- Air&Oxygen blender.
- Oxygen supply system.
- Baby cot is infinitely adjustable.
- 24 kinds of sound and light alarm, comprehensive protection of baby safety.
Technical Parameters:
| General Parameters | Item | Specifications |
| Power Supply | AC220V, 50/60Hz(Standard) 1320VA; AC110V, 60Hz(Optional) |
|
| Size(L*W*H) | 950*645*(1250~1450)mm | |
| Maximum Heater Output Power(220VAC) | Air Chamber Heater: 330W
Water Tank Heater: 155W |
|
| Auxiliary Network Power Output | AC220V/50Hz, both maximum permissible currents are 220V 1.5A | |
| Heater Indication | 0~100%, 10% increasing amount | |
| Temperature Control Mode | Bed temperature mode; The skin temperature mode | |
| Bed Temperature Control Temperature Range | 25.0~37.0℃;
37.1~39.0℃(Span mode) |
|
| Skin Temperature Control Range | 34.0~37.0℃; 37.1~38.0℃(Span mode) | |
| Skin Temperature Sensor Display Range | 5.0~65.0℃ | |
| Heating Time | ≤50min | |
| Temperature Variability | ≤0.5℃ | |
| Temperature Uniformity(Mattress in horizontal position) | ≤0.8℃ | |
| Temperature Uniformity(Mattress in tilted position) | ≤1.0℃ | |
| Difference Between Average Incubator Temperature and Control Temperature | ≤1.5℃ | |
| Skin Temperature Sensor Accuracy | ±0.3℃ | |
| Humidity Control | The running time of the tank after it is filled with distilled water | Humidity below 70%RH, at least 24h |
| Water Tank Capacity | 1500ml | |
| Humidity Display Range | 0%RH~99%RH | |
| Humidity Control Range | CAH%RH~90%RH(1% per adjustment) | |
| Humidity Control Accuracy | ±10% | |
| Oxygen Concentration Control | Oxygen Concentration Display Range | 0%~99% |
| Oxygen Concentration Display Resolution | 1% | |
| Oxygen Concentration Control Accuracy | ±5% oxygen volume concentration | |
| Oxygen Concentration Setting Range | 21%~60%(1% per adjustment) | |
| Oxygen Sensor Service Life | The maximum use time is 10000h at 100% oxygen concentration | |
| Transport Storage | Temperature | -20℃~+55℃ |
| Atmospheric Pressure Range | 500hPa~1060hPa | |
| Relative Humidity | ≤93%RH | |
| Packing Size/Gross Weight | 1150*800*1600mm
190kg |
|
| The Work Environment | Relative Humidity | 30%RH~75%RH |
| The Environment Temperature | +20℃~+30℃ | |
| Pressure Range | 700hPa~1060hPa | |
| Service Altitude | ≤2000m | |
| Over Voltage Category | II | |
| Pollution Levels | 2 | |
| Ambient Air Velocity | ≤0.3m/s | |
| Other Indicators | Environmental Noise In Incubator Baby Compartment | No more than 55dB(A) at stable temperature |
| Carbon Dioxide Concentration In The Infant Compartment | A mixture of 4% carbon dioxide and air was fed into an 8mm diameter tube in the vertical direction from the mat to the top of the box at 750mL/min to a point 10cm above the mattress center and 10cm above the center. The concentration was less than 0.5% after stabilization | |
| The velocity of air flow over a baby’s mattress | ≤0.35m/s | |
| General
Alarm |
Power failure alarm, fan alarm, sensor alarm, deviation alarm, over temperature alarm, water tank placement error alarm, water shortage alarm | |
| The System Alarm | Mcu1 and Mcu2 communication error alarm, Mcu2 and Mcu3 communication error alarm, analog-to-digital conversion chip 2544 failure alarm, analog-to-digital conversion chip 2548 failure alarm, over temperature alarm, temperature heating system failure, humidity heating system failure | |
Configuration:
| Model | MBI-B32 |
| Temp. Control | Bed/Skin Temperature |
| Display | LCD |
| Bed Temp. Control Range | 25℃~39℃ |
| Skin Temp. Control Range | 34℃~38℃ |
| Humidity Control Range | CAH%~99%RH |
| Oxygen Control Range | 21%~99% |
| Cabinet | Yes |
| Height Adjustable | Yes |