Contact

MPT-100

 

Blood Gas Analyzer

MPT-100

 

Blood Gas Analyzer:

Allinone Cartridge:

  • No fluidics in instrument, maintenance free
  • Selfcontained calibration, cleaning solution and biohazard waste collection pack
  • Optimized cartridge design minimizes clot formation

 

Easy Operation:

  • Simple 4step process, easy to use
  • Supports HL7 protocol for LIS/HIS
  • LCD display/Touch Screen for results
  • Large storage memory for minimum of 150 results or more
  • System should be capable of twopoint calibration along with manual calibration
  • Calibration history at least 24 hours
  • System should have at least 34 programs for quality control

 

High Efficiency:

  • Automatic calibration and continuous system performance monitoring
  • Built in printer, print results automatically after detection
  • Detailed Requirement:
  • Microprocessor based Blood gas Analyzer to measure, blood gas and electrolytes 33 parameters in only 60 seconds
  • Assay technology: Potentiometric and Amperometric electrodes (pH & O2 electrode), and a range of Ionselective electrodes.
  • Sample Type: Whole blood (syringe and capillary) with recommended anticoagulant.
  • Plasma/Serum for electrolytes only.
  • Easy changing access to electrodes and reagents.
  • Automatic sample recognition and sampling.
  • RS 232 or USB Interface

 

Ultimate Flexibility:

  • Compact size for various units (ICU, ED, OR, etc.)
  • Convenient handle, easy to cany
  • Builtin battery, 4 hours backup
  • Throughput: 30 samples/hour
  • Analysis Time: Not more than 80 seconds nominal
  • Control panel with easy setting and monitoring of pC02, p02, pH
  • Measuring range: (In Between)
  • pC02: 5 200 mmHg
  • p02: 0 800 mmHg
  • pH: 6.000 8.000 pH

 