Patient Cardiac Monitor

MRI-M12

Features

The monitor adopts Ergonomic design and a compact structure to meet the needs of various parameter detections. It is suitable for adults, children, and newborns.

  • Standard parameters: Sp02, 3/5 leads ECG, PR, HR, RESP, NIBP, 2*TEMP.
  • Optional parameters: EtC02, 12 leads ECG, 2*IBP.
  • 10/12/15 inch color LCD screen with optional touch screen and rotator knob.
  • Capable to visualize 6 or more waveform. Support color coding for different waveform.
  • 28 ST segment analysis with ST trend with continuous display of ST value on screen.
  • Multiple screen layout to view big font size in numeric and waveform.
  • Trending facilities for up to 360 hours of both graphical & numerical along with event review facility for all parameters and full waveform disclosure facility.
  • Different visual alarm with color coding and audible alarms for various parameters with three levels of volume adjustment.
  • Support HL7 protocol to access the central monitoring system, low-power design, extending battery life Configure, fanless design to ensure a quieter and cleaner environment Arrhythmia detection and analysis in various leads.

Product Performance

  • Standard application: 5-leads ECG, NIBP, Sp02, RESP, TEMP, PR/HR
  • Optional application: Touch screen, IBP, EtC02 (mainstream/sidestream), 12-lead ECG, recorder, CMS
  • Large font, dynamic short trend display interface, calculation of hemodynamics and cardiac output.

Specifications

Display

  • MRI-M12: 12 inches Color TFT LCD
  • Resolution: MRI-M121024*768 pixels

ECG

  • 3-lead: I, II, III
  • 5-lead: I, II, III, aVR, aVL, aVF, V
  • ECG Range: 0~100%
  • ECG Resolution: 2.36uV/LSB
  • HR Range: 15~300bpm(adult), 15~250bpm(pediatric/neonate)
  • HR Accuracy: ±1 bpm or ±1%, whichever is greater.
  • RR Range: 0~120bpm
  • RR Accuracy: 15~120rpm: ±2rpm or ±2% <15rpm: Undefined
  • Sweep speed: 6.25 mm/s, 12.5mm/s, 25mm/s, 50mm/s, less than ±5% error
  • Display sensitivity: 2.5 mm/mV (X0.25), 5 mm/mV (X0.5), 10 mm/mV (XI), 20 mm/mV (X2), 40
  • mm/mV (X4), Auto, less than ±5% error

SP02

  • SP02 Measurement Range: 0-100%
  • SP02 Accuracy: 70 to 100%: ±2% (adult/pediatric, without motion)
  • 70 to 100%: ±3% (neonate, without motion)
  • 0 to 69%: Not specified.
  • PI Range: 0-20%
  • PR Range: 25-250bpm
  • PR Accuracy: ±2bpm or ±2% (whichever is greater)

TEMP

  • Technique: Thermal resistance
  • Measurement Range: 0~50.0°C
  • Accuracy: ±0.2°C
  • Resolution: 0.1°C
  • Refresh Rate: Is
  • Alarm limit range: Temp High: (low limit +0.1°C) ~ 50.0°C
  • Temp Low: 0°C- (high limit -0.1°C)

NIBP

  • Measure method: Automatic oscillation method
  • Pressure Range: 0~300mmHg (0~40kPa)
  • Pressure Accuracy: ±2mmHg or ±1% of reading (whichever is greater)
  • Resolution: ImmHg or 0.1kPa
  • SYS Range: Adult: 40-270mmHg
  • Pediatric: 40-200mmHg
  • Neonate: 40-130mmHg
  • DIA Range: Adult: 10-210mmHg
  • Pediatric: 10-162mmHg
  • Neonate: 10-90mmHg
  • Mean Range: Adult: 20-230mmHg
  • Pediatric: 20-170mmHg
  • Neonate: 20-100mmHg
  • Accuracy: mean deviation<±5mmHg, standard deviation<8mmHg
  • Work pattern: manual, automatic, continuous measurement
  • Automatic mode measurement interval: lmin/ 2min/ 3min/ 4min/ 5min/ lOmin/ 15min/ 30min/ 60min/ 90min/ 2h/ 3h
  • Continuous mode measurement time: 5 min
  • Maximum single measurement time: 180s

Respiration

  • Technique: Trans-thoracic impedance
  • Lead: Options are lead I and II
  • Respiration excitation waveform: < 300pA RMS, 62.8kHz (±0%)
  • Respiration impedance range: 0.3-50
  • Baseline impedance range: 200-25000 (using an ECG cable with IkO resistance)
  • Bandwidth: 0.2~2Hz (-3 dB)
  • Sweep speed: 6.25 mm/s, 12.5mm/s, 25mm/s, 50mm/s

PR

  • Measurement range: 25bpm~250bpm
  • Resolution: lbpm
  • Alarm limit: PR High: (PR Low limit +lbpm)- 300bpm
  • PR Low: 15bpm- (PR High limit -lbpm)

HR

  • Measurement range: 15~300bpm (adult)
  • 15~350bpm (pediatric/neonate)
  • Resolution: lbpm
  • Accuracy: ±1 bpm or ±1%, whichever is greater
  • Sensitivity: 0.2mV

Data Storage

  • Trends: A minimum of 120 hours trend data with the resolution no less than 1 minute
  • Events: 200 events
  • NIBP measurements: 200 events
  • Interpretation of resting 12-lead ECG results: 20 sets

IBP (Optional)

  • Number of channels: 2
  • Pressure Range: -50~350mmHg (-6.7~40.0kPa)
  • Accuracy: ±2% or ±1 mmHg, whichever is greater (excluding sensor error)
  • Resolution: 1 mmHg (0.1kPa)
  • Refresh Rate: 1s
  • Sensitivity: 5pV/V/mmHg
  • Zero adjustment range: ±200 mmHg
  • Impedance range: 300-3000Q
  • Volume displacement: < 0.04mm3/100mmHg

EtC02 (Optional)

  • Measurement mode: Sidestream, Mainstream
  • Technique: Infrared absorption
  • Zero RR delay: 20s, 25s, 30s, 35s, 40s, 45s, 50s, 55s, 60s
  • C02 Range: 0-20.0 vol%
  • C02 Accuracy: 0-12 vol%: ± (0.2 vol%±2% of readings)
  • 12-20 vol%: ± (0.2 vol%±6% of readings
  • AwRR Measurement Range: 0~150rpm
  • Resolution: 1 rpm
  • AwRR Accuracy: Mainstream: 0~150rpm, ±lrpm
  • Sidestream: 0~69rpm, ±lrpm; 70~150rpm, Undefined
  • Alarm limit: EtC02 High: (EtC02 Low limit ±lmmHg)- 150mmHg
  • EtC02 Low: 0- (EtC02 High limit -ImmHg)
  • FiC02 High: 0- 150mmHg
  • AwRR High: (AwRR Low limit ±1 rpm)- 120rpm
  • AwRR Low: 0- (AwRR High limit -1 rpm)