Contact

MES-2000

Elisa Machine

MES-2000

 

Features

  • fully automated Microprocessor Controlled Micro Plate/Batch ELISA Analyzer with LCD Monitor and data recording system
  • Digital display for results
  • Unit with automatic sample identification, dilution, incubation, conjugation, washing and reading facility
  • programmable incubation and microplate shaking
  • work in sessions
  • 3 integrated liquid buffer tanks
  • multiple heads
  • LAN/ WLAN port to share the data with hospital network
  • RS 232 or USB Interface
  • Programmable for 2 Plates of 96 wells with filters 405620 nm with at least 6 filters
  • Throughput 2×96 samples test i.e. Infectious profile, Hormones Profile, Auto antibodies, VitD
  • level, Immunoglobulin, Tumor markers, Auto immune panel
  • 12 or more assays for microplate
  • Up to 12 assays profile
  • Unit based on at least 2 microplate principal and 2 incubation positions
  • System should have possibility to detect clot and to check volume intake
  • The facility for Auto clot detection, sensor probes for reagent and samples have been added
  • System should have possibility to run 24 or more controls with 18 or more position for reagents
  • 8 or more suction channels and at least 8 dispensing channel
Model MES-2000
Sample Unit Sample Racks Standard 192 sample positions(expandable, up to 552)
Pipetting 10~1000μl
Dispensing System 2 aspirating and dispensing probe( X-Y-Z movement)
Reagent Unit Reagent Position 23 reagent racks, editable
Pipetting Range 10~1000μl,  1μl stepping
Pipetting Precision (100μl)CV≤0.5%
Washing Unit Dispensing Time 4 minutes to whole 96 to microplates
Washing Probes Two line 8-nozzle manifolds
Wash Containers Cleaning fluid(15L), waste water(25L), Buffer 1(5L),  Buffer 2(2L),

Buffer 3(2L),  Buffer 4(2L), with liquid level-sensing, liquid shortage and full alarm

Waste Containers 25L with waste full sensor
Washing Residual <2μl
Reading Unit Reading Channels 8 independent photometric channels,  mono anddichromatic reading
Absorbance 0.0~3.0OD
Spectral Range 400~700nm
Optical Filters 2 standard filters(450, 630nm); 2 more on request (405nm/450nm/492nm/630nm)
Reading Accuracy ±1% from 0~1.5OD, ±2% from 1.5~3.0OD
Incubation Unit Incubators 4 independent units
Temperature Range RT to 45℃
Temperature Accuracy ±0.5℃
Data Management Software System Above Windows 7 system
LIS System Bi-direction, support HL7 protocol
Results Absorbance and results reviewable by software
Printer External optional
Working Conditions Power Supply AC220V±10%, 50/60Hz; 110V±10%, 60Hz, 400W
Temperature 10~30℃
Humidity 30%~80%
Size&Weight External Size (W*D*H) 1235*705*875mm
Net Weight 146kg
Package Size (W*D*H) 1380*860*1435mm
Gross Weight 220kg